Open original FDA record ↗Recall record
- FDA identifier
- Z-0560-2013
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Dec 19, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
GE Healthcare has became aware of two issues with the Advantage Workstation. 1) A possible mismatch between the label of the tracked vessel and the underlying image associated with the cardio- vascular applications of the Advantage workstation that may impact patient safety. 2) A refresh problem of the stenosis / aneurysm measurement tools associated with the cardio-vascular applications of
Source facts
- Action
- GE Healthcare sent two Urgent Medical Device Correction letters dated May 23, 2012, to all affected customers. Each letter addressed a separate issue. The letters described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. Customers were instructed to use the following precautions until corrected software was installed: 1. Please avoid the use of leading or trailing spaces in Vessel Names 2. Be sure to press "Show Tracking" after depositing new bifurcation points 3. When eating a multi-point vessel tracking, remove and replace the entire branch, not only the final end-point. 4. When analyzing multi-phase cardiac images, do not delete vessel branches from a phase other than the one in which it was created. Customers were instructed that the new application version that addresses this issue will soon be installed on all systems. Customers with questions or concerns were instructed to call the following: United States 1-800-437-1171. Japan 0120-055-919 For other countries they were instructed to contact their GE Healthcare Service Representative. For questions regarding this recalla call 262-513-4122
- Initiated
- 2012-05-29
- Root cause
- Software design
- Terminated
- 2013-01-17
- Distribution
- Worldwide Distribution - USA including AL, CA, CO, FL, GA, IL, IN, IA, MD, MA, MI, MN, MS, MO, MT, NE, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX , UT, VA, WI and Internationally to the following countries: VENEZUELA, UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, TUNISIA, THAILAND, TAIWAN, SWITZERLAND, SPAIN, SOUTH AFRICA, SINGAPORE, SERBIA, SAUDI ARABIA, RUSSIA, ROMANIA, REUNION, PORTUGAL, POLAND, PHILIPPINES, PERU, PANAMA, NORWAY, NIGERIA, NEW CALEDONIA, NEPAL, MEXICO, MALAYSIA, KOREA, JAPAN, ITALY, IRELAND, INDONESIA, INDIA, HONG KONG, GREECE, GERMANY, FRANCE, FINLAND, ESTONIA, EGYPT, ECUADOR, DENMARK, CZECH REPUBLIC, CHINA, CHILE, CANADA, BULGARIA, BRAZIL, BELGIUM, BELARUS, AUSTRIA, AUSTRALIA, ARGENTINA ALGERIA.
- Recall number
- Z-0560-2013
- Recall status
- Terminated
- CDRH record ID
- 108512
- FDA classified
- 2012-12-19
- Recall event number
- 61593