Skip to content

Public FDA record

Z-0561-06: BD Latitude (DMS) -- Blood Glucose Monitor.

Recall record Z-0561-06 for Becton Dickinson & Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0561-06
Company in FDA record
Becton Dickinson & Company
FDA classification date
Feb 25, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

BD Blood Glucose Monitors may inadvertently switch the Unit of Measure from mg/dL to mmol/L or vice versa during battery insertion or when a meter is dropped.

Source facts

Action
Recall notifications letters began mailing on 1/5/2006 to distributors, retailers, and DME providers. Healthcare professionals and consumers mailings will be during the week of 1/9/2006
Initiated
2006-01-05
Root cause
Other
Terminated
2006-09-05
Distribution
The firm distributes their blood glucose monitors Nationwide to distributors, retailers, durable Medical Equipment providers, Healthcare Professionals, and consumers. Foreign accounts include locations in Canada and Germany. There is one VA account.
Recall number
Z-0561-06
Recall status
Terminated
CDRH record ID
43538
FDA classified
2006-02-25
Recall event number
34292

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.