The system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the x-ray and gantry motions will both resume.
Source facts
Action
A letter dated 9/17/08 was sent to each account telling them that the PLC would be upgraded. The firm visited each account and made corrections.
Initiated
2008-09-17
Root cause
Device Design
Terminated
2009-12-23
Distribution
Nationwide Distribution
Recall number
Z-0561-2010
Recall status
Terminated
CDRH record ID
69359
FDA classified
2009-12-23
Recall event number
47370