Open original FDA record ↗Recall record
- FDA identifier
- Z-0561-2019
- Company in FDA record
- Helena Laboratories, Inc.
- FDA record creation date
- Dec 6, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to some cards in this lot reported to have experienced variations in results, (verified by wave form), resulting in some cards with shorter or longer times than expected.
Source facts
- Action
- On 02/24/2017, the firm sent an email to affected users, with a Withdrawal Notification. The firm informed the customers/distributors that Due to some cards in this lot experiencing variations in wave form, resulting in some cards with shorter or longer times than expected. The firm instructed the customers/distributors to disposes of affected lot, as a preventive quality measure. On 04/12/2017, the firm mailed formal letters to all customers. PUBLIC CONTACT: "Name: Dr. Jessica Jones Hanka "Title: Technical Services Manager "Address: 1530 Lindbergh Drive, Beaumont, TX 77707 "Telephone #: (409) 842-3714 x1177 "Email: jhanka@helena.com
- Initiated
- 2017-02-24
- Root cause
- Device Design
- Terminated
- 2023-10-20
- Distribution
- US: Products under control of Manufacturer in TX OUS: Four International Distributors - Australia, Netherlands, Italy, and England
- Recall number
- Z-0561-2019
- Recall status
- Terminated
- CDRH record ID
- 168423
- FDA record created
- 2018-12-06
- Recall event number
- 81378