Skip to content

Public FDA record

Z-0562-2013: Titanium Tack Starter Kit, REF 400-270, Rx Only, non-sterile, BIOHORIZONS, BIRMINGHAM, AL 35244. Dental

Recall record Z-0562-2013 for BioHorizons Implant Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0562-2013
Company in FDA record
BioHorizons Implant Systems Inc
FDA classification date
Dec 19, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

The cleaning instructions provided may not be sufficient to remove residual contaminants after surgical use.

Source facts

Action
BioHorizons sent an "URGENT DEVICE CORRECTION" letter on November 19, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers should contact the Customer Care Department at (888) 246-8338 for questions regarding this recall.
Initiated
2012-11-19
Root cause
Component design/selection
Terminated
2013-09-24
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Argentina, Australia, Brazil, Bulgaria, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Ecuador, Egypt, Fiji, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Lebanon, Lithuania, Malaysia, Mexico, Panama, Peru, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Spain, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.
Recall number
Z-0562-2013
Recall status
Terminated
CDRH record ID
114695
FDA classified
2012-12-19
Recall event number
63744

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗