Z-0562-2018: Ingenuity TF PET/CT (model 882442) running software version 4.0.2 This device is a diagnostic imaging s

Recall record Z-0562-2018 for Philips Medical Systems (Cleveland) Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0562-2018
Company in FDA record
Philips Medical Systems (Cleveland) Inc
FDA record creation date
Feb 9, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software issue causes PET reconstructions to fail intermittently. It was determined that reconstructions fail due to a negative table position (-1 value is inserted) in the raw data list file, rather than the actual table position. This error has been found to occur in two scenarios: 1) When the system operator cancels an acquisition a. The error will occur every time a scan is cancelled by the operator. 2) Couch position requests within the software sequence were delayed a. The error occurs intermittently, but has been found to occur more frequently when the gantry s network is heavily loaded with multiple retrospective reconstructions running in parallel. b. The error manifests to the technologist by an error message and Failed status on the Reconstruction Monitor and the error message Result {0} failed to reconstruct" on the Acquisition Workflow window during reconstruction after the patient scan has been completed. In both scenarios, the acquisition data will not be able to be reconstructed and will therefore be unusable.

Source facts

Action
Philips Healthcare sent an Urgent-Field Safety Notice dated December 11, 2017. U.S. Customers will be receive the letter via USPS Certified mail. Non-U.S.. customers will receive the letter through Philips Global Markets following local regulations. Philips is informing customers of the software issues on the systems, as well as available workarounds to prevent these issues from occurring or to minimize their impact. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, follow the prompts)
Initiated
2017-12-11
Root cause
Software design
Terminated
2022-08-30
Distribution
Worldwide Distribution - US Distribution to the states of : CA, OH, IL, PA and NY.
Recall number
Z-0562-2018
Recall status
Terminated
CDRH record ID
161075
FDA record created
2018-02-09
Recall event number
79008

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗