Z-0562-2022: Novalung Sensor Box (F30000163) a component of Novalung Console (F30000162)- Indicated for long-term (> 6

Recall record Z-0562-2022 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0562-2022
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA record creation date
Feb 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Error messages 206 (yellow) and 208 (red) technical failure, flow measurement during use of the Novalung System imply that the communication between flow sensor and sensor box is interrupted. An interrupted communication between flow sensor and sensor box might lead to a disabled flow measurement and air bubble detection.

Source facts

Action
Fresenius Medical Care issued Urgent Medical Device Correction Letter ( Novalung Sensor Box P/N F30000163 FA-2021-33-C) on 12/7/21 via USPS First Class Mail. Letter sates reason for recall, health risk and action to take: In the rare event 206/208 alarms occur, you may silence the alarms intermittently using the audio pause button. If you have continued alarms, it is recommended to replace either the sensor box or the entire console if a secondary or backup sensor box is available. Reference section 6.2 of the Instructions for Use (Part Number US (Part Number US-003-2019-08). If available, switch to a backup sensor box. To correct this issue, FMCRTG will update the sensor box on your Novalung console. An FMCRTG Field Service Technician will be contacting you to schedule this correction. For questions or concerns, please contact your local sales representative or FMCRTG Technical Support Services at 800-405-1321.
Initiated
2021-12-07
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, FL, IL, MI, NJ, NV, NY, OK, OR, PA, TN, TX, WV.
Recall number
Z-0562-2022
Recall status
Open, Classified
CDRH record ID
191361
FDA record created
2022-02-04
Recall event number
89369

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗