Open original FDA record ↗Recall record
- FDA identifier
- Z-0562-2022
- Company in FDA record
- Fresenius Medical Care Holdings, Inc.
- FDA record creation date
- Feb 4, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Error messages 206 (yellow) and 208 (red) technical failure, flow measurement during use of the Novalung System imply that the communication between flow sensor and sensor box is interrupted. An interrupted communication between flow sensor and sensor box might lead to a disabled flow measurement and air bubble detection.
Source facts
- Action
- Fresenius Medical Care issued Urgent Medical Device Correction Letter ( Novalung Sensor Box P/N F30000163 FA-2021-33-C) on 12/7/21 via USPS First Class Mail. Letter sates reason for recall, health risk and action to take: In the rare event 206/208 alarms occur, you may silence the alarms intermittently using the audio pause button. If you have continued alarms, it is recommended to replace either the sensor box or the entire console if a secondary or backup sensor box is available. Reference section 6.2 of the Instructions for Use (Part Number US (Part Number US-003-2019-08). If available, switch to a backup sensor box. To correct this issue, FMCRTG will update the sensor box on your Novalung console. An FMCRTG Field Service Technician will be contacting you to schedule this correction. For questions or concerns, please contact your local sales representative or FMCRTG Technical Support Services at 800-405-1321.
- Initiated
- 2021-12-07
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, FL, IL, MI, NJ, NV, NY, OK, OR, PA, TN, TX, WV.
- Recall number
- Z-0562-2022
- Recall status
- Open, Classified
- CDRH record ID
- 191361
- FDA record created
- 2022-02-04
- Recall event number
- 89369