Z-0563-2013: CoverStainer Slide Rack The CoverStainer Slide Rack is an accessory to the CoverStainer instrument. The

Recall record Z-0563-2013 for Dako Denmark A/S. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0563-2013
Company in FDA record
Dako Denmark A/S
FDA classification date
Dec 19, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

The expected life time of the CoverStainer slide racks (CS 10330) does not meet established performance requirements, originating from variations in the manufacturing process.

Source facts

Action
Dako sent a Voluntary Product Recall Notification dated November 13, 2012, to all affected customers. This voluntary recall notification was initiated because the expected life time of the CoverStainer slide racks (CS 10330) does not meet established performance requirements, originating from variations in the manufacturing process. The result can be that slides are falling out of the rack during the staining process. While this has been observed at only a few customer sites, the corrective action includes all CoverStainer installations. Customers were notified that Dako would correct the rack situation by using a new manufacturer and has imroved the rack design. Customers with questions were instructed to contact their local Dako representative. For questions regarding this recall call 805-566-3036.
Initiated
2012-11-26
Root cause
Device Design
Terminated
2013-05-14
Distribution
Worldwide Distribution - USA (nationwide) and the following countries: Australia France Rep. of Korea Belgium Germany Russia Brazil Israel Saudi Arabia Canada Italy Spain Chile Japan Sweden China Mexico Taiwan Denmark Netherlands Turkey Ecuador Norway UK Finland Poland
Recall number
Z-0563-2013
Recall status
Terminated
CDRH record ID
114796
FDA classified
2012-12-19
Recall event number
63807

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗