Z-0563-2021: Sentinel V10.x & V11.x, Models: 98200 & 98201 - Product Usage: designed using the latest programming tech

Recall record Z-0563-2021 for Spacelabs Healthcare, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0563-2021
Company in FDA record
Spacelabs Healthcare, Ltd.
FDA record creation date
Dec 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During system upgrade incorrect comments may be added to existing tests. In such instance: Resting ECG and ABP tests which did not contain comments at the time of the upgrade, may have unrelated comments added. Existing comments will not be overwritten or removed. Rhythm ECG Tests may have comments overwritten with unrelated comments. Any tests created after the upgrade are not affected.

Source facts

Action
On 11/06/20, Urgent Medical Device Correction notices were mailed to customers. Customers were asked to do the following: Any comments associated with ABP, Resting, or Rhythm tests that existed in your system prior to the date of your upgrade should be reviewed for accuracy before being used to inform any clinical decision. Weigh the benefits versus the risks when deciding whether to use the comments associated with ABP, Resting, or Rhythm tests which existed in your system prior to the date of your upgrade of Sentinel. We recommend that you brief your staff regarding the issue listed above. Complete and return the customer reply form or use the following link to acknowledge receipt of this letter: https://www.spacelabshealthcare.com/support/recall-reply-form/ For additional information or technical assistance, please contact: Global Technical Support at, +44 (0) 1992 507740, or gtsdc@spacelabs.com
Initiated
2020-11-06
Root cause
Software design
Terminated
2024-10-18
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of PR, MN, NE, FL, IL, CA, NY, ND, AR, WI, SD, TN, MO, MA, TX, MT, OH, MI, IA, NJ, CT, NC, UT, WV, GA, CO, PA, AL, DC, NV, DE, WA, OR, IN, SC, HI, MS, KS, ID, MD, LA, NH, VA, ME, AK, VT, OK, AZ, KY, NM, RI and the countries of AFGHANISTAN, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BELGIUM, BOLIVIA, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALY, KENYA, KOREA, REPUBLIC OF, KUWAIT, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, OMAN, PAKISTAN, PALESTINE, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, Republic of Macedonia, SAUDI ARABIA, Serbia, SINGAPORE, Slovakia, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VENEZUELA.
Recall number
Z-0563-2021
Recall status
Terminated
CDRH record ID
184456
FDA record created
2020-12-10
Recall event number
86753

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗