Z-0563-2022: 1. Arterial Line Kit 20ga x 6" 2. Arterial Line Kit /1 20ga x 6"

Recall record Z-0563-2022 for Argon Medical Devices, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0563-2022
Company in FDA record
Argon Medical Devices, Inc
FDA record creation date
Feb 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Guidewire included with two lots of Arterial Line Kits would not fit through the needle of the same kit. This could potentially lead to a slight delay in the procedure and minor blood loss.

Source facts

Action
Urgent Product Recall Notice was send out to the affected customers on 01/24/2022 via registered mail: The notification letter to each customer identify the specific lots and quantities that were shipped to that customer, and requesting to take the following action by customer. 1. Complete and return product inventory sheet quickly as possible 2. Notify/share the notification with anyone who needs to be aware within your organization or to any organization where the potentially affected products have been transferred. 3. Return all affected product to Argon Athens facility using RGA# provided in the notification. If you have any questions about the recall letter or about the recall action it describes in the notification, please contact Brian.Rogers@argonmedical.com. You may also contact Arbee.Cummings@argonmedical.com or Andrea.Wieczor@argonmedical.com
Initiated
2022-01-24
Root cause
Mixed-up of materials/components
Distribution
US Nationwide distribution in the states of AL, CA, DE, IL, IN, LA, MT, NV, OH, TX, VA, TN.
Recall number
Z-0563-2022
Recall status
Open, Classified
CDRH record ID
191587
FDA record created
2022-02-04
Recall event number
89439

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗