Open original FDA record ↗Recall record
- FDA identifier
- Z-0566-2007
- Company in FDA record
- Beckman Coulter Inc
- FDA classification date
- Mar 6, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There is a potential for erroneous yet credible results for Body Fluids on the LH700 series hematology analyzers when a cassette label fails to read and the subsequent sample is cycled in Body Fluid mode.
Source facts
- Action
- A Product Corrective Action (PCA) letter was mailed on Sep 12, 2006 to all LH 750 accounts to inform them that there is an issue with COULTER LH 700 Series analyzer software that could affect the Body Fluid analysis and that this this anomaly is present only in Software revisions of 2B. The letter instructs the customer on how to determine their software version. Immediate user preventive action was provided in the letter which is to reset the analyzer when a cassette label read error occurs. The letter advises that this information be shared with laboratory staff and be retained as part of their Quality System documentation. A customer response form was also included.
- Initiated
- 2006-09-12
- Root cause
- Other
- Terminated
- 2012-04-19
- Distribution
- Nationwide and Canada
- Recall number
- Z-0566-2007
- Recall status
- Terminated
- CDRH record ID
- 50283
- FDA classified
- 2007-03-06
- Recall event number
- 37287