Z-0566-2021: Microsurgical Operating Room Tables;7300B (18-350-70);

Recall record Z-0566-2021 for Mizuho America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0566-2021
Company in FDA record
Mizuho America, Inc.
FDA record creation date
Dec 14, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unintended tilting of the Tabletop MST 7300BX and incorrect operation due to infrared operating signal errors of Tabletops MST 7300B and MST 7300BX.

Source facts

Action
The firm's representatives notified their U.S. consignees regarding the recall notification by phone and a separate FedEx letter was sent (08/09/2019 - date the letter was issued). The Customer Notification Letter "IMPORTANT VOLUNTARY RECALL NOTIFICATION", informed customers of the two (2) types of errors: 1) Unintended Tilting of the Tabletop - Impurities in the hydraulic oil has caused minor leaking inside a control valve causing the table to tilt after remaining static for long periods of time. 2) Incorrect Operation due to Infrared Operating Signal Errors - The table can inadvertently mix infrared signals from the hand pendant while using in the wireless mode and may cause unintended movement. Customer are also being informed that a sales representative will contact them to organize the logistics of the correction for the tables. The recall notification includes to visit the firm's webpage (www.mizuho.com) to find additional information related to the recall. For additional questions, the Customer Service Group or local sales representatives can be contacted.
Initiated
2019-09-11
Root cause
Device Design
Distribution
US: AZ, CA, FL, MO, NC, OH, PA, TX, and WI. OUS: Greece, Israel, Japan, and Ukraine.
Recall number
Z-0566-2021
Recall status
Open, Classified
CDRH record ID
180895
FDA record created
2020-12-14
Recall event number
85442

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗