Open original FDA record ↗Recall record
- FDA identifier
- Z-0566-2024
- Company in FDA record
- CardioQuip, LLC
- FDA classification date
- Dec 15, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection.
Source facts
- Action
- On October 24, 2023, CardioQuip issued a "Urgent: Medical Device Correction/Removal" notification via USPS to all affected consignees. CardioQuip ask consignees to take the following actions: 1. CardioQuip recommends that you remove your potentially contaminated MCH device from service immediately. 2. CardioQuip Customer Service will contact you by November 3, 2023 to provide your facility with a loaner device and to schedule an IWPR procedure for your MCH device. 3. Please complete and return the Receipt of Communication/Response Form attached to this letter to CardioQuip as soon as possible. 4. Any adverse events experienced relating to CardioQuip devices should be reported to CardioQuip or through the FDA s MedWatch Program:
- Initiated
- 2023-10-24
- Root cause
- Process control
- Distribution
- US: California, Texas, Pennsylvania, Washington, Ohio, Nebraska OUS: None
- Recall number
- Z-0566-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204348
- FDA classified
- 2023-12-15
- Recall event number
- 93446