Z-0567-04: This device is labeled as Sterile and is packaged in a Tyvek pouch with the following labeling: PALMAZ Ge

Recall record Z-0567-04 for Cordis Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0567-04
Company in FDA record
Cordis Corporation
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A packaging defect may compromise the device's sterility barrier.

Source facts

Action
The firm initially mailed letters to accounts on 9/26/2003 instructing them to quarantine the indicated products pending a visit from Cordis inspectors. This mailing was followed up by one of two methods: 1) some accounts were visited by a Cordis Inspector who inspected the product at the account, and removed any potentially discrepant product and put a green dot on product which passed inspection and could be used, and 2) the remainder of the accounts were visited by Cordis Sales Representatives, in which case all product was removed from the account and returned to Cordis for inspection. Product passing Cordis inspection had a green dot put on the label and was redistributed. A representative of the account, and the inspector or sales representative signed an Acknowledgement form which in kept by Cordsi. This inspection process was completed by January 22, 2004.
Initiated
2003-09-26
Root cause
Other
Terminated
2009-10-05
Distribution
Product was distributed nationwide to 1035 domestic hospital accounts no international, government or military accounts.
Recall number
Z-0567-04
Recall status
Terminated
CDRH record ID
31725
FDA classified
2004-07-20
Recall event number
27450

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗