Open original FDA record ↗Recall record
- FDA identifier
- Z-0567-2021
- Company in FDA record
- Mizuho America, Inc.
- FDA record creation date
- Dec 14, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Unintended tilting of the Tabletop MST 7300BX and incorrect operation due to infrared operating signal errors of Tabletops MST 7300B and MST 7300BX.
Source facts
- Action
- The firm's representatives notified their U.S. consignees regarding the recall notification by phone and a separate FedEx letter was sent (08/09/2019 - date the letter was issued). The Customer Notification Letter "IMPORTANT VOLUNTARY RECALL NOTIFICATION", informed customers of the two (2) types of errors: 1) Unintended Tilting of the Tabletop - Impurities in the hydraulic oil has caused minor leaking inside a control valve causing the table to tilt after remaining static for long periods of time. 2) Incorrect Operation due to Infrared Operating Signal Errors - The table can inadvertently mix infrared signals from the hand pendant while using in the wireless mode and may cause unintended movement. Customer are also being informed that a sales representative will contact them to organize the logistics of the correction for the tables. The recall notification includes to visit the firm's webpage (www.mizuho.com) to find additional information related to the recall. For additional questions, the Customer Service Group or local sales representatives can be contacted.
- Initiated
- 2019-09-11
- Root cause
- Device Design
- Distribution
- US: AZ, CA, FL, MO, NC, OH, PA, TX, and WI. OUS: Greece, Israel, Japan, and Ukraine.
- Recall number
- Z-0567-2021
- Recall status
- Open, Classified
- CDRH record ID
- 180902
- FDA record created
- 2020-12-14
- Recall event number
- 85442