Z-0568-2021: Centeze Centesis Catheter, 6F, 10cm, 18Ga, 12cm, REF DRC-002-06 Centeze Centesis Catheter, 6F, 10cm, 18G

Recall record Z-0568-2021 for Galt Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0568-2021
Company in FDA record
Galt Medical Corporation
FDA classification date
Dec 14, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing issue, the catheter tip geometry is incorrect (thicker wall) resulting in a more obtuse tip angle.

Source facts

Action
On 11/16/2020, the firm sent an "URGENT-Medical Recall Notice" to customers to inform them that certain Lots of catheters may have a tip geometry with a thicker wall, resulting in a more obtuse tip angle. This obtuse tip angle, if present in the catheter tip, can cause increased insertion resistance through the epidermis layer of the skin. The Recalling Firm is instructing customers to: Note- An updated Customer Notification Letter was sent on 11/20/2020 via email. 1. Identify and segregate the affected lot(s) that are in their possession as well as those in the possession of their end-users. 2. Complete the enclosed Field Correction Reply Forms and email or fax them to the attention of Corrections Coordinator at quality@galtmedical.com or 214-778-1433. The Recalling Firm has enclosed one form to account for inventory at the customer's facility and an additional form for capturing the same details from their end-users. The form lists the product number, lot number and quantity that the customer has received. The Recalling Firm is pointing out that It is important that even if a customer does not have any product remaining in their possession that they fill out the attached forms (Attachment A & B) noting zero quantity to be returned and email or fax the forms to the Recalling Firm. 3. Ship the returned product to GALT MEDICAL CORP. using Galts carrier account information listed on the forms. 4. Reference Return Authorization Number RMA# GALT_2004 on the outside of the shipping box and include a copy of the Field Correction Reply Forms with your shipment. 5. Once the completed Field Correction Reply Forms have been received and processed, the Recalling Firm will issue a credit to the customer for the product returned. Customer are asked to place a new PO notating Replacement Parts for new products. The cause of the problem has been identified and corrective actions have been implemented to assure this problem does not reoccur. If you h
Initiated
2020-11-16
Root cause
Process control
Terminated
2022-11-07
Distribution
U.S.: FL, GA, MA, NC, NJ, OH, PA, Puerto Rico and UT. O.U.S: Canada, Netherlands, Turkey, United Kingdom, and UAE
Recall number
Z-0568-2021
Recall status
Terminated
CDRH record ID
184562
FDA classified
2020-12-14
Recall event number
86795

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗