Skip to content

Public FDA record

Z-0569-2015: 1488 HD 3-Chip Camera; Model Number 1488-210-105 - C-Mount. Product Usage: High definition camera to v

Recall record Z-0569-2015 for Stryker Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0569-2015
Company in FDA record
Stryker Endoscopy
FDA classification date
Dec 12, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In certain operating room environments, a colored screen image occurs with the 1488 Cameras that were manufactured during a specific time period

Source facts

Action
Stryker sent Device Recall letters to Risk Managers of affected sites on March 1, 2013. The issue was described and the fact that there have been no reports of patient harm. If user observe the colored screen prior to placement of the camera, unplug and reconnect the camera head into the console and/or power cycle the console to correct the issue. Local Stryker Representatives will coordinate the removal and replacement of the camera in the next several months. Questions may be directed to Stryker at 408-754-2517.
Initiated
2013-03-05
Root cause
Device Design
Terminated
2014-12-10
Distribution
Worldwide Distribution - US Nationwide in the states of Australia and Japan
Recall number
Z-0569-2015
Recall status
Terminated
CDRH record ID
131430
FDA classified
2014-12-12
Recall event number
69726

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗