Skip to content

Public FDA record

Z-0569-2021: DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral v

Recall record Z-0569-2021 for Ra Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0569-2021
Company in FDA record
Ra Medical Systems, Inc.
FDA record creation date
Dec 14, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.

Source facts

Action
On 10/29/2020, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Notification via certified mail informing them of the potential for the footswitch to fail to operate properly if liquid is spilled near or on the footswitch. On 11/16/2020, the firm sent out an updated customer notification. The firm is instructing customers to discontinue using the footswitch if liquids have been spilled on or splashed on the floor near or on the footswitch. The Recalling Firm will ship a replacement footswitch and will provide shipping materials for users to return the old footswitch. Instructions will be provided on how to install the new footswitch and return the old footswitch. Customers are asked to fill out and return the Acknowledgement and Receipt Form via FAX to 760-867-2706 or email at fieldcorrectiveaction@ramed.com. For any questions, contact the Recalling Firm at: 1-877-635-1800 Monday thru Friday 8:00 a.m. to 5:00 p.m. PST
Initiated
2020-10-16
Root cause
Device Design
Terminated
2023-01-11
Distribution
US Distribution to states of: AL, AZ, CA, FL, IL, IN, KY, LA, MD, NC, NV, NY, OK, OR, TX, and WA; and to OUS country: Japan.
Recall number
Z-0569-2021
Recall status
Terminated
CDRH record ID
184324
FDA record created
2020-12-14
Recall event number
86692

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗