Open original FDA record ↗Recall record
- FDA identifier
- Z-0569-2022
- Company in FDA record
- R82 A/S
- FDA record creation date
- Feb 8, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The chrome surface of the push brace can crack during use, potentially cutting the user.
Source facts
- Action
- A notification dated 12/10/2021 was sent to customers by R82 A/S notifying them of a Field Safety Notice concerning Push Brace 8910545 attached to patient mobility chairs. The notice communicates the chrome surface's potential for cracking during use, posing a risk of cutting the caregiver utilizing the push brace. The firm instructs customers to visually inspect recalled push braces before use and should be disposed of if any are detected. The firm also communicated that customers with affected push braces will receive replacements. If product has been further distributed, the notice needs to be forwarded on.
- Initiated
- 2021-12-08
- Root cause
- Process control
- Terminated
- 2023-09-05
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of North Carolina and California and the countries of Czech Republic, Finland, Germany, Greece, Holland, Italy, Japan, Norway, Poland, Russia, Sweden, and United Kingdom.
- Recall number
- Z-0569-2022
- Recall status
- Terminated
- CDRH record ID
- 190906
- FDA record created
- 2022-02-08
- Recall event number
- 89236