Open original FDA record ↗Recall record
- FDA identifier
- Z-0570-2008
- Company in FDA record
- BD Biosciences
- FDA classification date
- Mar 4, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Inaccurate results: Software error results in inaccurate display result statistics. If the user adjusts the lymph gate or attractors in the Lab Report view without using the Manual Gate function (as described in the system manual), the statistic results will not be updated to match the adjusted gate.
Source facts
- Action
- Becton Dickinson sent an Important Product Information letter by mail on 09/14/2007 informing customers of the need to follow the instructions for adjusting the lymph gate or attractors. The next software revision will alleviate this limitation.
- Initiated
- 2007-09-06
- Root cause
- Software design
- Terminated
- 2008-04-15
- Distribution
- Worldwide Distribution - USA, Belgium, Brazil, Peru, Singapore, Canada, Jamaica, Chile, Ecuador, Hong Kong-China, Thailand, Mexico, Venezuela, Barbados, Philippines, Puerto Rico, Guatemala, Japan, Uruguay, Argentina, China, Trinidad and Tobago, Honduras, Australia, New Zealand, and India.
- Recall number
- Z-0570-2008
- Recall status
- Terminated
- CDRH record ID
- 65479
- FDA classified
- 2008-03-04
- Recall event number
- 45447