Open original FDA record ↗Recall record
- FDA identifier
- Z-0570-2014
- Company in FDA record
- Philips Medical Systems, Inc.
- FDA classification date
- Dec 24, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
MRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHg
Source facts
- Action
- Philips Healthcare issued an Customer Information letter dated February 20, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer with the addendum to the Instructions for Use. If you have a ? in the parameter block and no CO2 waveform on the display, the waveform source is invalid. If invalid, check patient, confirm airway status and examine the FilterLine to make sure it is connected to the device and not kinked or pinched. For more information on the EtCO2 Option see Monitoring Carbon Dioxide chapter in the HeartStart MRx Instructions for Use. Should you have any questions or concerns, please contact your local Philips representative at 1-800-722-9377.
- Initiated
- 2013-12-10
- Root cause
- Software design
- Terminated
- 2017-04-18
- Distribution
- US Nationwide and OUS.
- Recall number
- Z-0570-2014
- Recall status
- Terminated
- CDRH record ID
- 124606
- FDA classified
- 2013-12-24
- Recall event number
- 67046