In certain operating room environments, a colored screen image occurs with the 1488 Cameras that were manufactured during a specific time period
Source facts
Action
Stryker sent Device Recall letters to Risk Managers of affected sites on March 1, 2013. The issue was described and the fact that there have been no reports of patient harm. If user observe the colored screen prior to placement of the camera, unplug and reconnect the camera head into the console and/or power cycle the console to correct the issue. Local Stryker Representatives will coordinate the removal and replacement of the camera in the next several months. Questions may be directed to Stryker at 408-754-2517.
Initiated
2013-03-05
Root cause
Device Design
Terminated
2014-12-10
Distribution
Worldwide Distribution - US Nationwide in the states of Australia and Japan
Recall number
Z-0570-2015
Recall status
Terminated
CDRH record ID
131431
FDA classified
2014-12-12
Recall event number
69726