Open original FDA record ↗Recall record
- FDA identifier
- Z-0570-2025
- Company in FDA record
- Securitas Healthcare LLC
- FDA classification date
- Nov 27, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain software versions will not notify end users of calls initiated by call points with the Arial Emergency and Nurse Call System.
Source facts
- Action
- An URGENT: MEDICAL DEVICE CORRECTION letter dated 10/31/24 was sent to customers. Necessary Actions to be Taken by Clients Instructions for determining which versions of the Arial Server Software you are using are attached hereto as Exhibit C. Clients using Arial Server Software versions 11.1.4.9, 11.1.5.5, or 11.1.6.2 should: 1. Not decline previously cleared calls through the Arial App; 2. Restart the Arial Server Software if an end user has declined a previously completed call using the Arial App. After restarting the Arial Server Software, calls from the affected call point(s) will begin to be received by the Arial Server Software and the Arial Server Software will notify end users as intended. Instructions for restarting the Arial Server Software are included on Exhibit B. 3. Contact Securitas Healthcare Technical Support at 1 (800) 380-8883 to upgrade the Arial Server Software to version 11.1.7 or later. For general questions concerning this communication please contact Securitas Healthcare Technical Support at 1 (800) 380-8883. We ask that you please complete and return a copy of the acknowledgement form attached hereto as Exhibit A.
- Initiated
- 2024-11-01
- Root cause
- Software Design Change
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Switzerland.
- Recall number
- Z-0570-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210892
- FDA classified
- 2024-11-27
- Recall event number
- 95652