Reason stated in the source

Incomplete upgrade: The software upgrade from February 1, 2007 may not have been completed. This software upgrade adds the algorithm (Slidetype) for the PATHWAY anti-HER-2/neu (4B5) antibody to the VIAS system.

Source facts

Action
Beginning in 6/07, Ventana Medical Systems, Inc., made phonecalls to labs asking to confirm software version, date of upgrade if available and confirmation of whether site uses 4B5 and Imaging together. A phone script was provided. Customers with systems not ugraded were to upgrade before further use of system. A notification letter dated 8/17/2007 was later sent to VIAS customers advising that the software upgrade from February 1 , 2007 may not have been completed on their system. This software upgrade adds the algorithm (Slidetype) for the PATWAY anti-HER-2/neu (4B5) antibody to their VIAS system . This Slidetype is cleared for marketing and is required prior to imaging slides stained with PATHWAY anti-HER-2/neu (4B5). The letter also advises that their VIAS and NEXES systems are conneded to the internet which enables this upgrade to be performed remotely with the support of the Ventana Medical Systems Customer Support Center (CSC). customers are instructed to call the CSC to carry out the upgrade or to schedule a Technical Applications Specialist (TAS) to assist with the upgrade in their lab. Arrangements for the upgrade are to be made prior to imaging slides stained with PATHWAY anti-HER-2/neu (4B5).
Initiated
2007-06-26
Root cause
Software change control
Terminated
2012-03-26
Distribution
Nationwide and Canada
Recall number
Z-0571-2008
Recall status
Terminated
CDRH record ID
66172
FDA classified
2008-03-18
Recall event number
45798

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗