Z-0571-2025: Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software

Recall record Z-0571-2025 for LivaNova Deutschland GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0571-2025
Company in FDA record
LivaNova Deutschland GmbH
FDA classification date
Nov 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.

Source facts

Action
On 10/18/24, a correction notice was emailed to customers: 1. Continue using the Heart-Lung Machine with SW HLM.1.5 as intended. 2. In the event of any cockpit reset during a procedure, use the backup control panel to make any setting adjustments as described in HLM IFU section 3.3.2 "Backup control panel." A. When prompted, press the "LAST CASE" button to retrieve the previous settings as described in HLM IFU chapter "7.1.3 Returning to the last case." If the gas blender settings were manually changed prior to the reset, adjust the settings as needed. B. If after selecting the "LAST CASE" button a second reset occurs, the "LAST CASE" button may not be available. Throughout the resetting of the system, the HLM's essential functions, including the pumps, alarms, sensors, and safety features, continue to operate as intended. After the second reset, the Gas Blender may go to Standby. If this occurs, the operator should reactivate the Gas Blender from the user interface on the Gas Blender unit to maintain operation of the Gas Blender for the procedure. After the second reset, a new case should be started from the home screen; while the machine distributes the profile again, the sensors are momentarily deactivated until the user selects the start case button. The user can skip the safety checks since they have already been performed before the reset and enter bypass mode. After starting the case, the user should update the case settings as needed. The start Case function is described in HLM IFU chapter "7.1.1 Starting a case" 3. Complete/return the response form to LivaNova.FSCA@livanova.com 4. Communicate the Correction to all personnel within your organization and any organization where affected devices have been transferred. Firm to contact customers to schedule a software update, available no later than 10/31/24. For questions regarding the correction, contact the firm at: LivaNova.FSCA@livanova.com.
Initiated
2024-10-18
Root cause
Software change control
Distribution
US Nationwide distribution in the states of CA, VA, GA, IN, NJ, FL, CO, MS, NV.
Recall number
Z-0571-2025
Recall status
Open, Classified
CDRH record ID
210913
FDA classified
2024-11-27
Recall event number
95665

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗