Skip to content

Public FDA record

Z-0571-2026: Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog Number: 455071P

Recall record Z-0571-2026 for Greiner Bio-One North America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0571-2026
Company in FDA record
Greiner Bio-One North America, Inc.
FDA classification date
Nov 19, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some of the Vacuette CAT Serum Separator Clot Activator tubes were manufactured without gel separator, which could yield erroneous assay results and require the patient to return to the lab for recollection or repeat testing. This may result in delayed diagnosis, cause a delay in clinical decision-making or treatment and in some rare cases, yield incorrect results from less stable parameters.

Source facts

Action
On 08/22/2025, the firm sent via email an "Urgent Medical Device Correction" letter to customer informing them that some Vacuette CAT Serum Separator Clot Activator tubes were manufactured without gel separator which may affect the stability of the sample. Customers are instructed to: " Stop use immediately: Discontinue use of the affected product (concerned Item/Lot) and isolate it within your facility. " Notify staff and customers: Inform all relevant employees and notify all customers to whom you distributed the affected products. " Assess risk: Evaluate potential risks or consequences in accordance with your internal procedures and take appropriate action. " Confirm product disposition: Complete the attached Product Disposition Site Confirmation Form and return it via fax to 877-217-7927 or email to recalls.us@gbo.com. Confirmation is required that all affected products have been discarded/destroyed. Replacement products will be provided upon receipt of your completed form. For questions, call Greiner Bio-One Technical Service at 800-515-8112.
Initiated
2025-08-19
Root cause
Labeling Change Control
Distribution
U.S.: AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, VA and WY. O.U.S.: None
Recall number
Z-0571-2026
Recall status
Open, Classified
CDRH record ID
215775
FDA classified
2025-11-19
Recall event number
97499

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗