Z-0572-2025: TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509)

Recall record Z-0572-2025 for Gauthier Biomedical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0572-2025
Company in FDA record
Gauthier Biomedical, Inc.
FDA classification date
Nov 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An out of specification component may result in the device exhibiting torque output higher than the design specification.

Source facts

Action
A Medical Device Recall (Removal) notification letter dated 11/11/24 was sent to the customer. Please Take the Following Steps: 1. Examine your inventory to determine if you have the affected lots and quarantine them immediately. DO NOT USE THE AFFECTED INSTRUMENTS. 2. Follow standard product return practices by contacting Heidi Wright at heidi.wright@gauthierbiomedical.com to issue an RMA number and coordinate the return of the affected product(s) to the following address: ATTN: Heidi Wright Gauthier Biomedical, Inc. 2221 Washington Street Grafton, WI 53024 Note: If the Verification Form is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual on this page of the notification. 3. Review, complete, sign and return the attached Verification Form to Heidi Wright at heidi.wright@gauthierbiomedical.com within five (5) business days of receipt of this notification. Please include in the email subject: CA 24-010. 4. Please complete the attached Verification Form even if you do not have the affected product lots on hand. 5. Forward this notice to anyone in your facility that needs to be informed (e.g., those who manage transport, store, stock, or use the affected products). 6. If any of the affected product lots have been forwarded to another facility, contact that facility and provide them with this notice. 7. Post a copy of this notice in a visible area for awareness and keep a copy for your records. 8. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA's MedWatch Adverse Event Reporting program online: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reportingprogram/reporting-serious-problems-fda. This medical device recall (removal) has been reported to the relevant health authorities. If you have any questions, please contact Heidi Wright, Quality Manager at Gauth
Initiated
2024-11-11
Root cause
Process change control
Distribution
US Nationwide distribution in the state of MA.
Recall number
Z-0572-2025
Recall status
Open, Classified
CDRH record ID
211094
FDA classified
2024-11-27
Recall event number
95746

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗