The parameters loaded in the analyzer for the hepatitis C virus do not match those in the package insert, and patient results may be reported out as false positives or false negatives.
Source facts
Action
Recall letters dated 2/15/05 are being sent to each consignee instructing them to discontinue use of pre-programmed HCV parameters with the Amplicor heatitis C virus test and that previous test results within a specific range be reviewed by a physician.
Initiated
2005-02-15
Root cause
Other
Terminated
2006-05-10
Distribution
Nationwide and Canada.
Recall number
Z-0574-05
Recall status
Terminated
CDRH record ID
37361
FDA classified
2005-03-08
Recall event number
31146