Open original FDA record ↗Recall record
- FDA identifier
- Z-0574-2020
- Company in FDA record
- Abbott Gmbh & Co. KG
- FDA classification date
- Nov 26, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a decrease in the control values and patient results. Patient values generated using the impacted calibrator lot 08219BE00 may have changed interpretation from reactive to falsely non reactive. Results impacted may present as a (falsely) non reactive result that fall in the range between 0.50 to 0.99 S/CO.
Source facts
- Action
- The firm initiated the recall by letter on 11/07/2019. The letter explained the issue and requested that the consignee discontinue use of the product and destroy it. The firm is providing replacement material.
- Initiated
- 2019-11-07
- Root cause
- Material/Component Contamination
- Terminated
- 2023-08-31
- Distribution
- US in the states of: AZ, CA, CO, CT, FL, KY, LA, MA, MD, MI, MN, MO, NJ, NY, OH, OR, RI, SC, TN, TX, WA, WV
- Recall number
- Z-0574-2020
- Recall status
- Terminated
- CDRH record ID
- 177818
- FDA classified
- 2019-11-26
- Recall event number
- 84238