Open original FDA record ↗Recall record
- FDA identifier
- Z-0575-2008
- Company in FDA record
- Bio-Rad Laboratories Inc
- FDA classification date
- Apr 22, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect values: An error was found in the BioPlex 2200 Software, Version 2.0 that results in assignment values being utilized for the BioPlex assay Calibrator Set that may be slightly different from those printed on the Value Assignment Data Sheet. Some values may be identical between the lot assignment combinations with the ANA Screen with MDSS kit.
Source facts
- Action
- Bio-Rad sent a Voluntary Field Correction notice dated November 1, 2007. Consignees were notified by phone call on Nov. 13, 2007, followed up by facsimile (domestic). Sub-recall to be by e-mail message for consignees outside the US. Distribution of the product ceased on Nov. 2, 2007. Bio-Rad will provide Calibrator Set CDs with only one lot of calibrator assignment values until a new version of software is available.
- Initiated
- 2007-11-13
- Root cause
- Software design
- Terminated
- 2008-07-23
- Distribution
- Worldwide Distribution - USA, Canada, France, and Germany.
- Recall number
- Z-0575-2008
- Recall status
- Terminated
- CDRH record ID
- 66439
- FDA classified
- 2008-04-22
- Recall event number
- 45892