Z-0575-2020: Alinity c Carbon Dioxide Reagent Kit, Produced for Abbott by Sekisui Diagnostics P.E.I. Inc. Prince Edwa

Recall record Z-0575-2020 for Abbott GmBH & Co. KG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0575-2020
Company in FDA record
Abbott GmBH & Co. KG
FDA record creation date
Nov 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Abbott internal studies have determined that atmospheric Carbon Dioxide (CO2) can be absorbed into the Alinity c Carbon Dioxide reagent, resulting in the potential for incorrect results. Abbott was made aware of this issue from customer complaints. Internal testing demonstrated that the amount of CO2 absorbed is higher with increased reagent carousel rotation and when the volume of reagent in the cartridge is reduced. This phenomenon can be detected as a shift in QC. The shift is predictive in both direction and magnitude of the potential impact to patient results.

Source facts

Action
On 10/23/19, a Product Recall letter was issued to all worldwide Alinity c Carbon Dioxide Reagent customers who received the in-date lots. US customers were sent this information via FedEx priority overnight. Customers are asked to immediately discontinue use and destroy product from LN 07P7230 lot 54449UQ03. For LN 07P7220 lot 54454UQ03, the product may continue to be used but 2 levels of CO2 controls must be run every hour until the new material is received. If you or any of the health care providers you serve have any questions regarding this information, US Customers can contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week). Customers outside the US , please contact your local area Customer Service.
Initiated
2019-10-23
Root cause
Under Investigation by firm
Terminated
2021-05-14
Distribution
Worldwide and US Nationwide in the states of: AL, AR, CA, CO, FL, GA, ID, IL, LA, MA, MI, MN, MO, NC, NY, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico Foreign: AUSTRALIA, BANGLADESH, BELGIUM, BOTSWANA, BRAZIL, CANADA, COLOMBIA, DENMARK, FRANCE, GREECE, Hong Kong, INDIA, ITALY, JAMAICA, LEBANON, MAURITIUS, NAMIBIA, NETHERLANDS, NEW ZEALAND, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, PORTUGAL, ROMANIA, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, UAE, and UNITED KINGDOM
Recall number
Z-0575-2020
Recall status
Terminated
CDRH record ID
177391
FDA record created
2019-11-26
Recall event number
84140

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗