- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Analyzers may fail to perform scheduled automatic two-point calibrations and Quality Control runs
Source facts
- Action
- Nova Biomedical Issued on 1/27/2006 an Advisory Notice to Stat Profile CCX Customers to report the malfunction and request the Good Laboratory Practices (GLP) be used and review be made to assure the calibration status and QC is being performed. A new software version is scheduled to be released to correct the error.
- Initiated
- 2006-01-27
- Root cause
- Other
- Terminated
- 2006-10-19
- Distribution
- Nationwide Foreign: Canada, UAE, Taiwan, Singapore, Budapest, UK, France, Malaysia, France, Mexico, Korea, Finland, Germany, Turkey, Ireland, Italy, Austria, Australia, Netherlands, Venezuela, Japan, Greece, Thailand, Israel, Spain, China, Hong Kong, Argentina, Czech Republic, Croatia,Trinidad WI