- FDA identifier
- Z-0576-2007
- Company in FDA record
- Medtronic Xomed, Inc.
- FDA classification date
- Mar 7, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A labeling discrepancy was internally identified for the surface electrode component of REF 8263210 AccuGuide EMG Injection Needle, 30G X 25MM. The outside box label correctly indicates 'Content: 5 sterile Injection Needles and 10 non-sterile Surface electrodes'; however, the pouches for the 10 surface electrodes inside the box are incorrectly labeled as 'sterile'.
Source facts
- Action
- Each consignee contacted via telephone 01/22/2007, and then sent a follow-up fax letter 02/02/2007, which explained the issue, and the options for dispositioning any product that they still had in stock. Since firm has determined that the risk of keeping and using these devices is negligible to no risk, as they are intended to be and typically used in a non-sterile manner, the firm is offering the consignees the option to keep their devices or have them replaced.
- Initiated
- 2007-01-22
- Root cause
- Other
- Terminated
- 2012-07-07
- Distribution
- Nationwide
- Recall number
- Z-0576-2007
- Recall status
- Terminated
- CDRH record ID
- 50302
- FDA classified
- 2007-03-07
- Recall event number
- 37298