Incorrect results: CRP assays may be calculated using incorrect calibration parameters which could result in falsely high or falsely low patient results being reported.
Source facts
Action
Roche Diagnostics contacted the consignee by phone on 12/4/07 and by follow-up letter dated 12/4/07. The user was instructed to discontinue use of two CRP assays on this analyzer. The letter described a workaround and that a new revision of software, due out first quarter 2008, will fix this.
Initiated
2007-12-04
Root cause
Software design
Terminated
2008-08-04
Distribution
USA Distribution in the state of Nebraska.
Recall number
Z-0576-2008
Recall status
Terminated
CDRH record ID
66617
FDA classified
2008-03-04
Recall event number
45984