Printouts may be printed in incorrect anatomical size when using syngo.plaza or syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of printer and the syngo.plaza printer configuration file. Printouts in anatomical size are not correct and may influence diagnostic decisions and/or therapy.
Source facts
Action
Siemens sent a Safety Advisory notices letter dated November 11, 2014 to affected customers. The letter identified the affected product, problem, and actions to be taken. For questions call 1-888-826-9702.
Initiated
2014-11-11
Root cause
Component design/selection
Terminated
2015-02-24
Distribution
US Nationwide Distribution
Recall number
Z-0576-2015
Recall status
Terminated
CDRH record ID
131566
FDA classified
2014-12-15
Recall event number
69802