Open original FDA record ↗Recall record
- FDA identifier
- Z-0577-2010
- Company in FDA record
- Ellex Medical Pty Ltd
- FDA classification date
- Jan 20, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Potential for two screws important in the design of the safety filter to come loose.
Source facts
- Action
- Ellex issued a bulletin stating the problem and identifying the affected products. Customers are to remove the Motorised Safety Filter from clinical use if the green laser light is visible while viewing through the microscope at any time during use of the Solitaire laser. The Bulletin also states that customers should contact an Ellex approved Service Engineer to investigate and perform the Field Corrective Action listed. The manufacturer will remedy the defect or bring the product into compliance free of charge. Ellex will inspect and rework all affected safety filters sold prior to February 2009.
- Initiated
- 2009-02-11
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2012-09-07
- Distribution
- Worldwide Distribution.
- Recall number
- Z-0577-2010
- Recall status
- Terminated
- CDRH record ID
- 87209
- FDA classified
- 2010-01-20
- Recall event number
- 54019