Z-0577-2015: Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service

Recall record Z-0577-2015 for Invacare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0577-2015
Company in FDA record
Invacare Corporation
FDA classification date
Dec 12, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.

Source facts

Action
Invacare sent an Urgent - Medical Device Recall letter to customers, dated October 2014. The letter identified the affected product as well as the problem involved. The letter instructed customers to examine their device and confirm that it has an affected serial number, or if service was conducted on the device using IRC5O2V Service Part Number 1165099 with the affected date codes. Customers were asked to immediately contact their dealer or service provider at the number listed, so that a replacement can provided. A service provider or dealer should be contacted with questions regarding the instructions.
Initiated
2014-09-23
Root cause
Process control
Terminated
2015-06-18
Distribution
Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.
Recall number
Z-0577-2015
Recall status
Terminated
CDRH record ID
130540
FDA classified
2014-12-12
Recall event number
69441

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗