Open original FDA record ↗Recall record
- FDA identifier
- Z-0577-2015
- Company in FDA record
- Invacare Corporation
- FDA classification date
- Dec 12, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.
Source facts
- Action
- Invacare sent an Urgent - Medical Device Recall letter to customers, dated October 2014. The letter identified the affected product as well as the problem involved. The letter instructed customers to examine their device and confirm that it has an affected serial number, or if service was conducted on the device using IRC5O2V Service Part Number 1165099 with the affected date codes. Customers were asked to immediately contact their dealer or service provider at the number listed, so that a replacement can provided. A service provider or dealer should be contacted with questions regarding the instructions.
- Initiated
- 2014-09-23
- Root cause
- Process control
- Terminated
- 2015-06-18
- Distribution
- Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.
- Recall number
- Z-0577-2015
- Recall status
- Terminated
- CDRH record ID
- 130540
- FDA classified
- 2014-12-12
- Recall event number
- 69441