Open original FDA record ↗Recall record
- FDA identifier
- Z-0577-2020
- Company in FDA record
- Boston Scientific Corporation
- FDA record creation date
- Nov 26, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process.
Source facts
- Action
- The firm, Boston Scientific, sent an "Urgent Medical Device Removal" letter dated 10/28/2019 overnight mail delivery to its customers. The letter stated the issue, identified affected product and stated that product should not be used. The customers were instructed, if any product is identified to segregate immediately and return to Boston Scientific. The customers were also instructed to Immediately complete the RECALL REMOVAL REPLY VERIFICATION TRACKING FORM and Scan/e-mail to: BSCFieldActionCenter@bsci.com OR Fax to #: 1-866-213-1806. This form must be completed and returned in all cases even if you do not have any affected product. If you are a distributor, please note that the removal notification should be forwarded to your customers. If you have any questions, contact Boston Scientific Quality Systems at 763-494-1133 or email: BSCFieldActionCenter@bsci.com.
- Initiated
- 2019-10-24
- Root cause
- Process control
- Terminated
- 2023-03-20
- Distribution
- Nationwide including Washington D.C. and Puerto Rico. OUS to include Canada.
- Recall number
- Z-0577-2020
- Recall status
- Terminated
- CDRH record ID
- 177422
- FDA record created
- 2019-11-26
- Recall event number
- 84152