The secondary supplier provided material was within specification but did not meet our design needs compared to the primary manufacturer. The product is designed for patient transfer, not lifting. However, we recognize that the product may be utilized outside of its intended design scope, and lifting may occur. Lifting could lead tearing of the equipment cover and the potential for patient falling through.
Source facts
Action
The firm initiated their recall by email on 12/27/2017 to their direct consignees. The consignees were directed to report back with the number of products on hand for return and to notify their customers of the recall.
Initiated
2017-12-27
Root cause
Device Design
Terminated
2018-07-30
Distribution
TX, CT, TN, MO, WI, FL, SD, MS, CA, MS
Recall number
Z-0578-2018
Recall status
Terminated
CDRH record ID
160825
FDA classified
2018-01-09
Recall event number
78899