Open original FDA record ↗Recall record
- FDA identifier
- Z-0579-2018
- Company in FDA record
- GE Medical Systems China Co., Ltd.
- FDA record creation date
- Feb 9, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Possible incomplete seal between the CO2 absorber and the breathing circuit CO2 bypass port assembly of the Carestation 600 Series systems.
Source facts
- Action
- On approximately 10/11/2017 customers were notified of the recall via letter. Instructions include that if elevated FiCO2 is observed, to increase the flow of fresh gas to reduce the volume of patient gas that is rebreathed. If the FiCO2 cannot be adequately reduced with this action, consider switching to another anesthesia delivery device. GE Healthcare is planning to release revised parts that minimize the likelihood of incomplete gas flow through the CO2 absorbent canister. GE Healthcare will correct all affected devices by installing updated parts at no cost to the customer. The updated parts will be included in the 12-month preventative maintenance schedule for the Carestation 600 Series System.
- Initiated
- 2017-10-11
- Root cause
- Other
- Distribution
- Distributed domestically to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, VT, WA, WI, WV. Distributed internationally to Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote Di'Viore, Croatia, Czech Republic, Democratic Republic of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Malta, Mauritius, Mexico, Moldova, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, Zimbabwe.
- Recall number
- Z-0579-2018
- Recall status
- Open, Classified
- CDRH record ID
- 159630
- FDA record created
- 2018-02-09
- Recall event number
- 78469