Skip to content

Public FDA record

Z-0579-2020: NC Emerge Monorail PTCA Dilatation Catheter. Sterilized using ethylene oxide. Packaged in the following

Recall record Z-0579-2020 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0579-2020
Company in FDA record
Boston Scientific Corporation
FDA record creation date
Nov 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process.

Source facts

Action
The firm, Boston Scientific, sent an "Urgent Medical Device Removal" letter dated 10/28/2019 overnight mail delivery to its customers. The letter stated the issue, identified affected product and stated that product should not be used. The customers were instructed, if any product is identified to segregate immediately and return to Boston Scientific. The customers were also instructed to Immediately complete the RECALL REMOVAL REPLY VERIFICATION TRACKING FORM and Scan/e-mail to: BSCFieldActionCenter@bsci.com OR Fax to #: 1-866-213-1806. This form must be completed and returned in all cases even if you do not have any affected product. If you are a distributor, please note that the removal notification should be forwarded to your customers. If you have any questions, contact Boston Scientific Quality Systems at 763-494-1133 or email: BSCFieldActionCenter@bsci.com.
Initiated
2019-10-24
Root cause
Process control
Terminated
2023-03-20
Distribution
Nationwide including Washington D.C. and Puerto Rico. OUS to include Canada.
Recall number
Z-0579-2020
Recall status
Terminated
CDRH record ID
177427
FDA record created
2019-11-26
Recall event number
84152

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗