- Recall classification
- Class II
Reason stated in the source
Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.
Source facts
- Action
- On October 14, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: Carefulle read the notification and ensure that all personnel are completely knowledgeable and thoroughly trained on the content of the letter. Complete the required training provided in the letter. Keep a copy of the notification with the operation manual. It is recommended to receive on-site TJF duodenoscope reprocessing training at your facility. If you have further distributed this product, identify and forward them this notification.
- Initiated
- 2025-10-14
- Root cause
- Device Design
- Distribution
- US Nationwide distribution.