Open original FDA record ↗Recall record
- FDA identifier
- Z-0580-2008
- Company in FDA record
- Siemens Medical Solutions Diagnostics
- FDA classification date
- Jun 5, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect results: Falsely elevated Troponin values have been reported which are inconsistent with the patient's clinical picture and test negative by other Troponin assays.
Source facts
- Action
- Two Support Bulletins (078N0361-01 & 078N0361-02) were sent by e-mail to all affected Siemens Diagnostics Branches to inform them of this issue. The first one was sent on 6/07/07 and the second one was sent on 7/05/07. Two Customer Bulletins (078D0647-01 & 078D0651-01) were sent by e-mail to the Siemens Diagnostics Branches for communication with the affected customers. The first one was sent on 6/07/07 and the second one was sent on 7/05/07. These bulletins state that if a false positive is suspected, the sample should be diluted and retested.
- Initiated
- 2007-06-07
- Root cause
- Material/Component Contamination
- Terminated
- 2012-08-28
- Distribution
- Worldwide Distribution - USA, Argentina, Austria, Australia, Brazil, Canada, Chile, Colombia, Egypt, France, Germany, Greece, Hong Kong, India, Israel, Italy, Malaysia, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Venezuela
- Recall number
- Z-0580-2008
- Recall status
- Terminated
- CDRH record ID
- 64326
- FDA classified
- 2008-06-05
- Recall event number
- 44725