Skip to content

Public FDA record

Z-0580-2014: steroEOS Workstation with software versions: 1.3, 1.4 and 1.5 used in processing 3D X-ray images Product

Recall record Z-0580-2014 for EOS Imaging. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0580-2014
Company in FDA record
EOS Imaging
FDA classification date
Dec 26, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Error copying information to the clinical file and to the report when two 3D models are opened simultaneously on the sterEOS workstation

Source facts

Action
EOS imaging issued a safety information letter dated November 29, 2013, to the US customer sites. The letter identifed the affected product, problem and actions to be taken. Users of the sterEOS workstation are asked not to simultaneously open several pairs of images in sterEOS 3D. The letter informed customers that EOS imaging will, at it's own expense, corect all systems affected by means of a software update. An EOS imaging representative will contact affected customers to make an appointment for this update once it is available. Users of the sterEOS workstation are asked not to simultaneously open several pairs of images in sterEOS 3D. For questions call +33 (0) 155 25 6060.
Initiated
2013-11-29
Root cause
Software design
Terminated
2015-01-13
Distribution
USA Nationwide Distribution
Recall number
Z-0580-2014
Recall status
Terminated
CDRH record ID
124401
FDA classified
2013-12-26
Recall event number
66980

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗