Open original FDA record ↗Recall record
- FDA identifier
- Z-0580-2014
- Company in FDA record
- EOS Imaging
- FDA classification date
- Dec 26, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Error copying information to the clinical file and to the report when two 3D models are opened simultaneously on the sterEOS workstation
Source facts
- Action
- EOS imaging issued a safety information letter dated November 29, 2013, to the US customer sites. The letter identifed the affected product, problem and actions to be taken. Users of the sterEOS workstation are asked not to simultaneously open several pairs of images in sterEOS 3D. The letter informed customers that EOS imaging will, at it's own expense, corect all systems affected by means of a software update. An EOS imaging representative will contact affected customers to make an appointment for this update once it is available. Users of the sterEOS workstation are asked not to simultaneously open several pairs of images in sterEOS 3D. For questions call +33 (0) 155 25 6060.
- Initiated
- 2013-11-29
- Root cause
- Software design
- Terminated
- 2015-01-13
- Distribution
- USA Nationwide Distribution
- Recall number
- Z-0580-2014
- Recall status
- Terminated
- CDRH record ID
- 124401
- FDA classified
- 2013-12-26
- Recall event number
- 66980