Z-0580-2018: Carestation 620 A2, Carestation 650 A2, and Carestation 650c A2

Recall record Z-0580-2018 for GE Medical Systems China Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0580-2018
Company in FDA record
GE Medical Systems China Co., Ltd.
FDA record creation date
Feb 9, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Possible incomplete seal between the CO2 absorber and the breathing circuit CO2 bypass port assembly of the Carestation 600 Series systems.

Source facts

Action
On approximately 10/11/2017 customers were notified of the recall via letter. Instructions include that if elevated FiCO2 is observed, to increase the flow of fresh gas to reduce the volume of patient gas that is rebreathed. If the FiCO2 cannot be adequately reduced with this action, consider switching to another anesthesia delivery device. GE Healthcare is planning to release revised parts that minimize the likelihood of incomplete gas flow through the CO2 absorbent canister. GE Healthcare will correct all affected devices by installing updated parts at no cost to the customer. The updated parts will be included in the 12-month preventative maintenance schedule for the Carestation 600 Series System.
Initiated
2017-10-11
Root cause
Other
Distribution
Distributed domestically to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, VT, WA, WI, WV. Distributed internationally to Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote Di'Viore, Croatia, Czech Republic, Democratic Republic of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Malta, Mauritius, Mexico, Moldova, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, Zimbabwe.
Recall number
Z-0580-2018
Recall status
Open, Classified
CDRH record ID
159631
FDA record created
2018-02-09
Recall event number
78469

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗