Open original FDA record ↗Recall record
- FDA identifier
- Z-0580-2025
- Company in FDA record
- Micro-X Ltd.
- FDA classification date
- Nov 27, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
Source facts
- Action
- On 10/22/24, Service Bulletin SB007-23-1.0 was emailed to customers restating receiving, storage, and operation conditions and informing customers that extreme humidity environments can adversely affect the equipment. For further information contact your local service representative: phone: 1-206-249-8764 Email: support-us@micro-x.com
- Initiated
- 2024-10-22
- Root cause
- Environmental control
- Distribution
- US: TX, PR, MD, CA
- Recall number
- Z-0580-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210817
- FDA classified
- 2024-11-27
- Recall event number
- 95624