Open original FDA record ↗Recall record
- FDA identifier
- Z-0580-2026
- Company in FDA record
- Medtronic Sofamor Danek USA Inc
- FDA classification date
- Nov 21, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.
Source facts
- Action
- Medtronic Sofamor Danek notified consignees on about 09/16/2025 via letter. Consignees were informed of updates to the IFU and surgical guide. Consignees were instructed to share the notification within each facility with clinical users of the affected systems and to complete and return the provided Customer Acknowledgement form.
- Initiated
- 2025-09-16
- Root cause
- Labeling design
- Distribution
- Worldwide distribution - US Nationwide and Puerto Rico. The countries of Australia, Austria, Belgium, Canada, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Netherlands, New Caledonia, New Zealand, Norway, Portugal, Reunion, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
- Recall number
- Z-0580-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216383
- FDA classified
- 2025-11-21
- Recall event number
- 97714