Reason stated in the source
Error messages, "Wash Aspirator Failure" and "Wash station unable to recover" are not generated when one or both HM wash probes are totally occluded.
Source facts
- Action
- The recalling firm notified consignees by telephone to inform them of the problem. A follow-up letter, "Urgent Field Safety Notice" dated 9/30/08 was issued with information describing the problem, the possible impact of an occluded wash probe on assay results, and emphasizing regular performance of routine wash probe maintenance as detailed in the operator guide to prevent wash probe occlusion. For questions, please contact the Siemens Technical Solution Center at 800-441-9250.
- Initiated
- 2008-09-30
- Root cause
- Software design
- Terminated
- 2009-05-05
- Distribution
- Nationwide Distribution including states of AR, CA, CO, IA, ID, IL, IN, LA, MD, MI, MO, MS, NC, NE, OH, OK, PA, SD, TN, TX, VA, and WI.
- Recall number
- Z-0581-2009
- Recall status
- Terminated
- CDRH record ID
- 73915
- FDA classified
- 2009-01-13
- Recall event number
- 49608