Z-0582-2019: Medtronic HeartWare Ventricular Assist HVAD System Battery Charger Model 1600US - Product Usage: The Hear

Recall record Z-0582-2019 for Heartware, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0582-2019
Company in FDA record
Heartware, Inc.
FDA record creation date
Dec 7, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

HVAD Battery Charger units manufactured with wrong inductors.

Source facts

Action
On November 2018 Medtronic sent the following letter to all their consignees: Dear Physician or Healthcare Professional: HeartWare, now a part of Medtronic, has been notified by an outside vendor that a quantity of HVAD battery charger units (see image above) may have been manufactured with an incorrect circuit component. This subset may exhibit one or more of the battery charging bays not fully charging the battery and/or battery charger indicators (LED) not lighting. The behaviors may occur in any individual bay, independent of the other charging bays. In analyzed units, up to 2 charging bays exhibited one of the above behaviors. In all cases, at least 2 of the remaining bays were able to continue to charge batteries as needed. We are not able to predict if, or when, one of the identified battery charger units may malfunction. The potential exists, however, that all circuits responsible for charging batteries and LED function may fail due to the incorrect component. In the population of potentially affected product, Medtronic has observed a malfunction rate of 16.7%. The number of potentially affected battery chargers remaining in distribution is estimated to be 2031 worldwide. These units were distributed between June 2017 and June 2018. Charging units outside of the serial numbers listed in Appendix A, all new manufactured battery chargers, and the individual batteries are not in scope of this advisory. Through November 02, 2018, Medtronic has received zero (0) reports of patient injuries associated with this issue. Patient Management Recommendations We realize that each patient requires unique clinical considerations. In consultation with Medtronics Independent Practitioner Quality Panel (IPQP), the following recommendations are provided: For patients with a HeartWare Battery Charging Unit in the identified list of Serial Numbers (refer to Appendix A): " Instruct patients to monitor the battery charging unit for any of the following beha
Initiated
2018-11-19
Root cause
Other
Terminated
2020-07-21
Distribution
Worldwide distribution - US nationwide distribution in the state of AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI. Countries of Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Italy, Latvia, Lebanon, Netherlands, Poland, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom.
Recall number
Z-0582-2019
Recall status
Terminated
CDRH record ID
169057
FDA record created
2018-12-07
Recall event number
81609

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗