Skip to content

Public FDA record

Z-0582-2020: PowerPort isp M.R.I. Implantable Port with Attachable 8F Polyurethane Open-Ended Single-Lumen Venous Cath

Recall record Z-0582-2020 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0582-2020
Company in FDA record
Bard Peripheral Vascular Inc
FDA record creation date
Nov 27, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaints received that products packaged with the incorrect introducer sheath size.

Source facts

Action
On 2/14/2019, the firm sent an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" via FedEx informing them the firm has identified one product code/lot number (RECS2407) may be at risk of incorrectly containing a 6.5Fr introducer sheath instead of an 8Fr introducer sheath. Customer are instructed to: 1. Do not use or further distribute any affected product. 2. Check all inventory locations within their institution for affected product code / lot number combinations listed in the recall notice. If you have further distributed any of the product code / lot numbers, and to immediately contact that location, advise them of the recall, forward these instructions and have them return the affected product to the Recalling Firm. 3. Remove any identified product from their shelves. 4. If the customers have used the affected product, to complete and return the attached Recall and Effectiveness Check Form indicating no product will be returned. Once the product affected by this recall has been removed from customer inventory: 5. Fill out the Recall and Effectiveness Check Form and to be sure to state the quantities and lot numbers of each recalled product that is intended to be returned. It is extremely important that the Recalling Firm receive this information. 6. Please call Customer Support Center at 1-800-321-4254 Option #5 (M-F 6am to 3pm MST) or email at BPV.CustomerSupportCenter@crbard.com. Once all information has been verified, Customer Support Center will issue you a Return Authorization (XC) Number to facilitate the expedient return of the product. The Recalling Firm will issue a replacement product for its customer's returned product. 7. Email the completed Recall and Effectiveness Check Form to BPV.CustomerSupportCenter@crbard.com or fax it to BPV at 1-800-994-6772. If you cannot email or FAX the form, please call Customer Support Center at 1-800-321-4254 Option #5 and report the required information verbally. 8. A mailing label will be enclosed for the custo
Initiated
2019-02-14
Root cause
Process control
Distribution
US - Nationwide OUS - None
Recall number
Z-0582-2020
Recall status
Open, Classified
CDRH record ID
177460
FDA record created
2019-11-27
Recall event number
84169

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗